PiLas Control Units EIG1000D and EIG1000AF

Parameter EIG1000D
(for PiLas 1 MHz version)

EIG1000AF
(for PiLas 100 MHz version)

                                                   internal trigger

repetition rate 1 MHz max 100 MHz fixed
other repetition rates between 10 and 110 MHz on request
repetition rate of internal trigger adjustable at 1 MHz, 500 kHz, 200 kHz,
100 kHz…10 Hz, single shot
100 MHz fixed
other repetition rates between 10 and 110 MHz on request

                                                  external trigger

external trigger input single shot – 1 MHz typ. ± 5% of the internally fixed repetition rate
amplitude -5V…+5V / TTL / NIM -5V…+5V / TTL / NIM
impedance 1 MOhm / 50 Ohm 50 Ohm
pulse width typ. 10 ns typ. 4 ns bei 100 MHz

                                                 pre-trigger output

amplitude TTL TTL or 2V @ 50 Ohm
pulse width typ. 10 ns typ. 4 ns bei 100 MHz
trigger delay
(others on request)
70 ns app. 60 ns
typical jitter between pre-trigger and optical output typ. 3 ps typ. 4 ps
ALS GmbH reserves the right to change specifications without prior notice.

Dimensions of EIG: width x height x depth = 235mm x 88mm x 326 mm

top home

Advanced Laser Diode Systems A.L.S. GmbH
Schwarzschildstr. 6, D-12489 Berlin
Phone/Fax + 49 (0) 30 6392 652- 0/1